ETS-Lindgren HI-4422 Field Probe (Archived) User Manual

Page 112

Advertising
background image


Probe

Description

Probe Response

Command

Dx

Read probe data.

• x=1, enables short

form output

• x=2, enables long form

output

For D1 command:

Dxx.xxuuu, the short form

output

• xx.xx=the reading, and the

position of the decimal point

depends upon the range

setting of the probe

• uuu=units

• _V_ = V/m, mW2 =

mW/cm², _V2 = [V/m]²

(underscore indicates a space)

For D2 command:

Dxx.xxuuurrrobaaa, the long

for output

• xx.xx=the reading, as

described for D1

• uuu=units, as described for

D1

• rrr=recorder out value, a

three-digit ASCII number

from 0 to 255

• o=over range indicator:

N=ok, O=over range

• b=battery status: N=safe

operating level, W=warning

level, F=fail level

• aaa=axis enable:

E=enabled, D=disabled,

axis order=X, Y, Z

112

|

Operating Protocols

Archived 6/29/10

Use only the content for

HI-4422 in this manual.

Disregard content for

all other probe models;

see ETS-Lindgren

website for other

probe content.