Probe
Description
Probe Response
Command
Dx
Read probe data.
• x=1, enables short
form output
• x=2, enables long form
output
For D1 command:
Dxx.xxuuu, the short form
output
• xx.xx=the reading, and the
position of the decimal point
depends upon the range
setting of the probe
• uuu=units
• _V_ = V/m, mW2 =
mW/cm², _V2 = [V/m]²
(underscore indicates a space)
For D2 command:
Dxx.xxuuurrrobaaa, the long
for output
• xx.xx=the reading, as
described for D1
• uuu=units, as described for
D1
• rrr=recorder out value, a
three-digit ASCII number
from 0 to 255
• o=over range indicator:
N=ok, O=over range
• b=battery status: N=safe
operating level, W=warning
level, F=fail level
• aaa=axis enable:
E=enabled, D=disabled,
axis order=X, Y, Z
112
|