Probe

Description

Probe Response

Command

TF

Read temperature in

Fahrenheit

:Txxxx.<CR>

<

Null

><

CR

>

Send the ASCII null

character

:N<CR>

<Null><CR> is a special command that can be used as the initial

command to the probe after it is turned on.

HI-4422

F

IELD

P

ROBE

When remotely operating the probe with commands manually entered from a

computer keyboard, the HI-4422 may sleep between commands, especially if the

sleep timer is set to a one or two second interval. The HI-4422 may appear

unresponsive to commands, or may appear to skip commands. If this occurs,

send an S0 command, and keep sending the command as fast as possible until

the probe responds with :S, and the sleep timer is turned off. Then the probe will

provide the proper responses.

Probe

Command

Description

Probe Response

Axxx Axis

enable/disable.

x=E (enable) or D

(disable)

xxx order=X axis, Y

axis, Z axis

B

Read battery voltage.

Bxx.xx,

xx.xx=battery voltage

Operating Protocols

|