KEYENCE DL-RS1A/IG User Manual

Page 16

Advertising
background image

14

Parameters of Commands and Responses

115

Sub display's screen

*

1

R/W

0: R.V. value screen
1: Analog output screen
2: HIGH setting value screen
3: LOW setting value screen
4: Shift target value setting screen
5: Calculated value display screen

0

116

Tolerance tuning Tolerance
setting width*

9

**.***

6

R/W

00.000 to 99.999

00.100

117

Calibration function SET1

±**.***

7

R/W

-99.999 to +99.999

+00.000

118

Calibration function SET2

±**.***

7

R/W

-99.999 to +99.999

+05.000

119

Calibration Function

*

1

R/W

0: Standard
1: User setting

0

120

System parameter*

10

**

2

R/W

0 to 15

00

129

Calculation function*

11

*

1

R/W

0: The calculation function is not used.
1: Addition mode
2: Subtraction mode
3: 2 Heads mode

0

130

Measurement mode

*

1

R/W

0: Edge control/Positioning mode
1: Outer diameter/Width measurement mode
2: Inner diameter/Opening measurement mode
3: Glass edge mode
4: Pin position measurement mode
5: Pin interval judgment mode
6: Pin diameter judgment mode
7: Specified edges interval measurement mode

0

131

Measurement direction

*

1

R/W

0: Top
1: Bottom

0

132

Averaging*

12

**

2

R/W

0: hsp
1: 1
2: 2
3: 4
4: 8
5: 16
6: 32
7: 64
8: 128
9: 256
10: 512
11: 1024
12: 2048
13: 4096

05

133

Output mode

*

1

R/W

0: NO (Normally Open)
1: NC (Normally Close)

0

134

Hold function setting

*

1

R/W

0: Sample hold
1: Peak hold
2: Bottom hold
3: Peak-to-peak hold
4: Auto peak hold
5: Auto bottom hold

0

135

Auto peak hold or Auto bottom
hold Trigger level*

7

±**.***

7

R/W

-99.999 to +99.999

+00.100

136

Timing Input

*

1

R/W

0: Level
1: Edge

0

137

Delay Timer

*

1

R/W

0: Delay timer off
1: On delay timer
2: Off delay timer
3: One shot timer

0

138

Timer value

****

4

R/W

1 to 9999

0060

139

Hysteresis*

9

**.***

6

R/W

0 to 99.999

00.010

140

Analog Output Scaling

*

1

R/W

0: Initial state
1: Free range

0

141

Analog output Lower limit*

7

±**.***

7

R/W

-99.999 to +99.999

+00.000

142

Analog output Upper limit*

7

±**.***

7

R/W

-99.999 to +99.999

+10.000

143

External input*

13

*

1

R/W

0: Initial state
1: User setting

0

Data

No.

Data name

Data

format*

1

No. of

bytes

Attribute*

2

Data range

Default value

Parameters of Commands and Responses

*1

In the Data format column, "±" indicates that the value can be either "+" or "-" and "*" signifies a
number from "0 to 9".

*2

Indicates that the data can only be read from (R), can only be written to (W), or can be both read
from and written to (R/W) the sensor amplifiers.

*3

The command is executed only when the data is changed from 0 to 1. To re-execute the
command, first change the setting to 0, then change it to 1. Otherwise, the command is not
executed.
If the sensor amplifier is turned OFF after the standard waveform registration is performed, the
previous state before the standard waveform registration is performed is restored. If you wish to
keep the standard waveform registration even after the power is turned OFF, set the saving the
standard waveform function to ON.
If the sensor amplifier is turned OFF after the zero shift is performed, the previous state before
the zero shift is performed is restored. If you wish to keep the zero shift state even after the
power is turned OFF, set the saving zero shift value function to ON.

144

External input 1*

13

*

1

R/W

0: Zero shift input
1: Bank A input
2: Bank B input
3: Laser emission stop input
4: Not use

0

145

External input 2*

13

*

1

R/W

0: Reset input
1: Bank A input
2: Bank B input
3: Laser emission stop input
4: Not use

0

146

External input 3*

13

*

1

R/W

0: Timing input
1: Bank A input
2: Bank B input
3: Laser emission stop input
4: Not use

0

147

External input 4*

13

*

1

R/W

0: Gain input
1: Bank A input
2: Bank B input
3: Laser emission stop input
4: Not use

4

148

Saving the standard waveform
function

*

1

R/W

0: OFF
1: ON

0

149

Saving zero shift value function

*

1

R/W

0: OFF
1: ON

0

150

Interference prevention
function*

14

*

1

R/W

0: OFF
1: ON

0

151

Display Digit

*

1

R/W

1: 0.001
2: 0.01
3: 0.1
4: 1

2

152

Power save function

*

1

R/W

0: OFF
1: Half
3: All

0

153

Position monitor

*

1

R/W

0: Initial state
1: OK/NG display
2: Red OFF
3: OFF

0

154

Display color

*

1

R/W

0: GO Green
1: GO Red
2: Always Red

0

155

Edge check function Number of edges

**

2

R/W

0-99

01

156

Edge check function

*

1

R/W

0: OFF
1: Setting A
2: Setting B

0

Data

No.

Data name

Data

format*

1

No. of

bytes

Attribute*

2

Data range

Default value

Advertising