Settings for data saving, Data storage function – KEYENCE TM-3000 Series User Manual
Page 12
12
VII -
11
Manual
P5-98
Obtain measurement data. Up to 65536 of measurement value can be stored for each OUT number.
Data storage function
XXX
Storing measurement data
Screen 1
Setting procedure
Settings for Data Saving
Settings for Data Saving
This section explains the settings for saving and loading measurement results.
VII
Screen 2
Select “Common” on the setting screen (screen 1).
Press the [ENTER] key on the console to select the item.
Select “Storage”.
Select “Data storage” from the pull-down menu.
Slide the [PROG/RUN] switch on the side of the console to move to screen 2.
On screen 2, press the [MENU] key on the console to display “Display setting” dialog.
Press the “Storage start/stop” button to start storage.
Screen 1
Data storage timing
Data storage timing using the data storage function is
set via “Data out timing” in “Common”.
Save timing can be selected from the following two.
•
Image update timing : Data is input each time the image is
updated. (Image update is performed for
each trigger input.)
•
TIMING 1
: Data is stored each time TIMNG 1
signal is input.
Slide the [PROG/RUN] switch on the side of the console to display the settings screen (screen 1).
The “Storage” mark is
displayed when storage
begins.
1
2
3
4
5
1
2
3
5
1
2
3