Using the pulley jig – KEYENCE LS-9030D User Manual

Page 2

Advertising
background image

2

E LS-9006D/LS-9030D

z

LS-9030D

To use LS-9030D’s air purge unit, prepare the following off-the-shelf items:

• PC6-M5 (PISCO fitting) (Recommended)
• Air tube Outer diameter

φ

6

1

Remove the set screw from the inlet to which the fitting will be
attached. (2 mounting points)

Recommended tightening torque: 1 to 1.5 N m

2

Insert the air tube into the fitting.

When cleaning the receiver, remove the air purge unit that is secured

with four screws as shown below.
Maximum tightening torque for securing the unit: 1.5 N m

Using the Pulley Jig

The following describes how to mount the pulley unit to the casing. The

compatible combinations for pulley models and sensor heads are shown

below.

LS-9006D : OP-87750

LS-9030D : OP-87749

For LS-9006D

1

Place the jig into the notched section of air purge unit.

2

Secure it using the provided hexagon socket bolts (M5 x L15).

For LS-9030D

1

Place the jig into the notched section of air purge unit.

2

Secure it using the provided hexagon socket bolts (M5 x L15).

If you use an L-shaped fitting, make sure that the threaded
portion will not become loose due to twisting of the tube.

• Supply air flow rate:

40L/min to 80L/min (20 L/min to 40L/min per unit)

• Air pressure:

Approx. 40 kPa.

The air pressure is measured before the air purge. The
pressure will vary depending on the point of measurement.
Be sure to check the performance based on the air flow.

• Be sure to use clean, dry air.
• The air purge performance and the sizer’s measurement

accuracy can be affected by the usage environment and/or
the condition of the air source. Determine the specifications
for the air source and the glass cleaning schedule in
accordance with the actual operating conditions.

NOTICE

Important

Advertising
This manual is related to the following products: