4-2 During Use

To level the surface of the sample; adjust the leveling feet so that the center of the narrow part

of the right and left sensor plates is on the liquid surface.

The viscosity of a liquid is temperature dependent and changes by negative 2 to negative 10

percent, per degree Celsius. Take changes in the liquid temperature into consideration for an

accurate measurement.

Be sure to calibrate using the standard viscosity fluid or purified water before measurement.

In a measurement that takes a long time, perform calibration periodically, as necessary.

Placing the sensor plates and the temperature sensor in the sample may change the sample

temperature. For precise measurement, leave the sample as is for a while, after placing the

sensor plates and the temperature sensor, to ensure no changes to the sample temperature.

And then, start a measurement.

Ensure a stable power source when using the AC adapter.

Use only your finger to press the keys. Using a sharp instrument such as a pen may damage

keys.

The sample cup is made of polycarbonate (PC) and is not appropriate for organic solvents.

When organic solvents are used as a sample fluid, do not use the accessory sample cup. Use

the glass sample cup (AX-SV-35) that is sold separately or a commercially-available glass

beaker.

The protector can be raised or removed. So, even when a beaker is used, the viscosity can be

measured with a small amount of sample.

How to remove the protector:

Press the left and right side frames lightly in the

direction indicated as 1 to remove the rotational axis.

Pull the protector in the direction indicated as 2 to

remove.