Data Output Format (type)

The output format appropriate for the device connected to RS-232C can be selected.

Parameter

Settings Description

0

A&D standard format

Used with the printer MODE 1 or MODE 2 when the optional compact
printer, AD-8121B is connected. Only the viscosity value is output.

1

• D.P. format

Used with the printer MODE 3 when the optional compact
printer, AD-8121B is connected.

With "prt 0" or "prt 1" selected for "Data output mode (prt)",
output contents can be selected by the settings of "5-at",
"5-td" and "5-ed".

With "prt 2" selected for "Data output mode (prt)", only the
viscosity value is output.

2

CSV format

Appropriate when a personal computer is used to collect data.

Measurement values are output in comma separated format.

With "5-td" and "id" settings, the date/time and ID number

can be added to the measurement data.

When a comma is selected as the decimal point by "pnt 1", a

semicolon ";" is used as a data separator.

The viscosity value and the temperature are output using the
internal resolution. *3

3

RsVisco format

Used with the graphing program RsVisco.
When a measurement is started using RsVisco, the viscometer
automatically selects this format.
The viscosity value and the temperature are output using the
internal resolution.*3

*3 The relation between the measuring unit and the internal resolution is as follows:

Viscosity Temperature

Model

mPa・s Pa・s cP P

°C

°F

SV-10 0.01 0.0001 0.01 0.0001

Internal
resolution SV-100

- 0.01 - 0.1

0.01 0.01

Measurement Elapsed Time Output (5-at)

Parameter

Settings Description

0

No output

1

• Output

With D.P. format selected, whether or not to add the
measurement elapsed time (the time elapsed from a
measurement start) to the measurement data can be selected.

For examples of output format, refer to "7-4 Data Output
Format Examples".