A&D Weighing SV-100 User Manual

Page 28

Advertising
background image

26

6-2-3 Simplified Calibration Using Purified Water (SV-10 only)

1 Place the purified water in the sample cup.

2 Press the START key to measure

the purified water. Confirm that the
viscosity and temperature values are
stabilized.

3 Press and hold the START key.

The theoretical viscosity

value(Viscosity

× Density)

of the

purified water at the measuring
temperature is displayed and all the
displays blink.


To cancel the operation, press the

STOP key. The display returns to

the status before calibration.


4 Press the START key again to

perform calibration.
When calibration is completed, " end "
is displayed. Then, the viscosity
returns to the measurement mode.