A&D Weighing SV-100 User Manual

Page 48

Advertising
background image

46

Output format example (2) With date and time added

Function setting

(

√=Output Blank=No output)

5-td 1 Date/time

5-ed 0

Device

ID number

SV-10 output format example

Viscosity / Temperature

Display

Output format example

mPa

⋅s /

°C

1.23 mPa

⋅s

,2003/03/19,12:34:56,+025.67,C,+00001.23,mPa s

: Space (ASC 20h)

SV-100 output format example

Viscosity / Temperature

Display

Output format example

Pa

⋅s /

°C

1.23 Pa

⋅s

,2003/03/19,12:34:56,+025.67,C,+00001.23, Pa s

: Space (ASC 20h)

Output format example (3) To output the measured temperature and viscosity only

Function setting

(

√=Output Blank=No output)

5-td 0 Date/time

5-ed 0

Device

ID number

SV-10 output format example

Viscosity / Temperature

Display

Output format example

mPa

⋅s /

°C

1.23 mPa

⋅s

,,,+025.67,C,+00001.23,mPa s

: Space (ASC 20h)

SV-100 output format example

Viscosity / Temperature

Display

Output format example

Pa

⋅s /

°C

1.23 Pa

⋅s

,,,+025.67,C,+00001.23, Pa s

: Space (ASC 20h)

Outputs in the order of date, time, temperature,
temperature unit, viscosity and viscosity unit.
The output data are 46 characters long excluding the
terminator.

Outputs in the order of temperature, temperature unit,
viscosity and viscosity unit.
The output data are 28 characters long excluding the
terminator.