A&D Weighing SV-100 User Manual

Page 61

Advertising
background image

59

*1 When a sample cup of 45 mL is used.

*2 Repetitive

measurement

with the sensor plates remaining in the sample

*3 The value after calibration using a standard viscosity fluid at a temperature range

between 20°C and 30°C with no condensation.

In a measurement that takes a long time, perform calibration using a standard viscosity
fluid or purified water periodically, as necessary.

*4 The unit switches to Pa

⋅s.

*5 The operating temperature of each standard and optional accessory is as follows:

Sample cup (Standard and optional)

AX-SV-33:

0 to 120°C

Small sample cup (Optional)

AX-SV-34:

0 to 120°C

Glass sample cup (Optional)

AX-SV-35:

0 to 230°C

Water jacket assembly (Optional)

AX-SV-37:

0 to 100°C

Glass storage container (Optional)

AX-SV-38:

0 to 180°C (Lid: 80°C max.)

Take the operating temperature shown above into considerations when using the above
items.