A&D Weighing SV-100 User Manual

Page 25

Advertising
background image

23

6-2-1 One-point Calibration

1 In the standby mode, press and hold

the HOLD key to enter the
calibration mode. " Cal " appears.

2 Select one-point calibration (Cal-1)

and press the PRINT key to confirm.
The standby mode of the one-
calibration mode appears.

Use the MODE key to switch
between the calibration modes,
one-point calibration (Cal-1) or
two-point calibration (Cal-2).
Press the STOP key to exit the
calibration mode. The display returns
to the standby mode.

3 Place the standard viscosity fluid in

the sample cup. Press the START
key to start a measurement.

4 After the measurement, wait for the

display to become stable and press
the PRINT key. The measurement
value blinks and is ready to be
corrected.

5 Correct the value using the following keys:

MODE key

Switches the blinking digits.

START key

Changes the setting of a
blinking digit.

STOP key

Moves the decimal point.

6 Press the PRINT key to confirm the correction value.

7 To exit the calibration mode:

With " end " being displayed, press the PRINT
key again to return to the standby mode.

To correct the value:
With " end " being displayed, press the

STOP key and correct the value.

8 Measure the viscosity of the standard

viscosity fluid used. Confirm that the
viscometer displays the similar value
as the entered correction value.

The below is an example of the SV-10.
For the SV-100, the unit is Pa

⋅s

(Standby mode)

(One-point calibration)

Press and hold

(One-point calibration

standby mode)

(Measurement mode)

When the data

is stabilized

(Correct the value)

(End of one-point calibration)

(Standby mode)

To correct

Set standard viscosity fluid

(Confirm the value)

Advertising
This manual is related to the following products: