12. TROUBLESHOOTING

The viscometer is a precision instrument. When the measuring environment or the measurement
method is inadequate, correct values can not be obtained. If measurement values do not become
stable or they seem to be incorrect, check as described below. If improper performance persists after
checking, contact the local A&D dealer for repair.

When measurement values do not become stable:

Is the ambient environment free from vibration and drafts?

Places such as second or higher floor or near busy highways or rail lines are prone to

vibration.
Avoid these places or use an anti-vibration table, AD-1685.

Reconsider the setting of "Condition (Cond) ” of the function setting.

Set it to "Cond 2".

Avoid direct drafts in the vicinity of the viscometer.

Is there a strong electrical or magnetic noise source such as a motor near the viscometer?

Install the viscometer away from the electrical or magnetic noise sources.

Is the protector or the sensor protective cover in contact with the sensor plates or the

temperature sensor?

Attach the protector and the sensor protective cover properly so that they do not touch

the sensor plates or the temperature sensor.

Remove the protector, the surface locator plate or the sensor protective cover when

necessary.

How to remove the protector
















Press the left and right side frames lightly in the direction indicated as 1 to remove the

rotational axis. Pull the protector in the direction indicated as 2 to remove.