How to remove and attach the surface locator plate

Removing

Loosen the screw and remove the surface

locator plate from the temperature sensor.

Attaching

Install the surface locator plate as

shown in the illustration. Slip the locator

plate onto the temperature sensor.

Move the plate to the top of the sensor.

Adjust the position of the locator

plate so that the tip of the locator

plate aligns to the center of the

narrow part of the sensor plates.

Tighten the screw.

Note
When the sensor plate and the surface locator plate are too close, a measurement
error may occur due to the liquid surface tension. Therefore, secure a clearance of
at least 1 mm between the sensor plate and the surface locator plate. Maintain the
clearance by rotating the surface locator plate when necessary.


How to remove the sensor protective cover

(Sensor unit side view)

Remove the protector and the surface locator plate.

Pull the sensor protective cover in the direction indicated as 1 to release the portions A

(two) and B from the protector frame.

Pull the sensor protective cover in the direction indicated as 2 to remove.

Note
When removing the sensor protective cover, it may touch the sensor plates and the
temperature sensor. Use much care not to damage them.