A&D Weighing SV-100 User Manual

Page 26

Advertising
background image

24

6-2-2 Two-point Calibration

1 In the standby mode, press and hold

the HOLD key to enter the
calibration mode. " Cal " appears.

2 Select two-point calibration (Cal-2)

and press the PRINT key to confirm.
The standby mode of the two-
calibration mode appears.

Use the MODE key to switch
between the calibration modes,
one-point calibration (Cal-1) or
two-point calibration (Cal-2).
Press the STOP key to exit the
calibration mode. The display returns
to the standby mode.

3 In two-point calibration mode, the calibration

mode ID indicator (-) appears below the
temperature display.

4 Place the standard viscosity fluid in the

sample cup. Press the START key to start
the measurement of the first point.

5 After the measurement, wait for the display

to become stable and press the PRINT key.
The measurement value blinks and is ready
to be corrected.

6 Correct the value using the following keys:

MODE key

Switches the blinking digits.

START key

Changes the setting of a
blinking digit.

STOP key

Moves the decimal point.

7 Press the PRINT key to confirm the correction value.

8 To correct the value:

In the calibration standby mode to enter the
second point, press the STOP key and
correct the value.

The below is an example of the SV-10.
For the SV-100, the unit is Pa

⋅s