LABEC M-TIT-5 User Manual
Page 20
 
7.1.1 mV/pH Titration Parameter
mV/pH titration parameter is used to indicate the titration is mV titration or pH titration.
When user select mV titration, the current titration is mV titration, all displayed parameter are 
according to mV and vice versa. 
7.1.2 Balance Potential Parameter
The process of pre-titration mode is: Firstly meter will add certain amount of titrant, when 
adding finish, meter sampling the potential value, then calculate the amount to add for next 
time according to potential changing value between before adding and after adding. Adding 
again, sampling, calculating and repeat like this till titration finished. After titrant added, 
meter will sampling the potential after adding and calculate the amount to add for next time, 
so the potential sampling is very important. The meter allow user to set potential balance 
range which is also called potential changing range. When titration started, the meter will 
judge, sampling current potential, if the change of current potential satisfy the balance 
potential set by user, this potential is valuable, otherwise the meter will keep on waiting till 
the potential is satisfied. Of cause, meter allows user to switch off the judgment of balance 
potential. 
 
7.1.3 Balance Time Parameter
Same as balance potential parameter, balance time parameter is also used to control the 
potential sampling after adding titrant. When potential varies big and cannot satisfy the 
balance condition, it is necessary to use time parameter to limit, otherwise the meter will 
always wait for the balance of potential. The meter allows user to set balance time i.e. if 
potential cannot balance, the meter can still continue titration when set balance time is up. So 
if potential satisfy balance potential condition first, balance time condition will be invalid; 
when user switch off balance potential parameter, the meter will take balance time parameter 
as valuable (user is not able to switch off Balance Time Parameter and Balance Potential 
Parameter as the same time); vice versa. 
 
7.1.4 Minimal Volume Parameter
User is allowed to set minimal adding volume, usually is 0.02mL 
 
7.1.5 End Point Jump Value Parameter
End point jump value is the final basis for meter to judge the end point, it is divided into three 
grades: large, middle and small, user only need to select large/middle/small and no need to set 
specific jump value. If user observes that end point jump of titration is somewhat low or noise 
is too large, with the result that titration end point is difficult to find, we suggest that user 
reset this part of parameter (set end point jump to "small" if end point jump is low, and set 
end point jump to "middle" or "large" if noise is large). 
In order to satisfy the requirement of multi-end point titration, the meter allows five end 
points at most in any titration. There are five independent controlled values corresponding to 
five end points. The controlled value can be set to be large or small independently. It is 
impossible that various titration to have the united mode or method. In consideration of 
demand of general users, we divide end point jump into three grades that is large end point 
19