LABEC M-TIT-5 User Manual

Page 38

Advertising
background image

9.1 Generation of Mode

When titration finished, if necessary, user could generate titration mode, please see chapter
8.1.2 Generation of Mode for details.


9.2 Viewing Mode

User could view and amend relative parameter by viewing the mode generated by himself or the
system mode provided by meter. In starting status, press ‘View’ in menu, select ‘Viewing Mode’
and press ‘Enter’ to view the mode. Displayed as follows:

Mode HC1 → NaOH (0.1)

pH Titration

Number of end point: 1
2003/06/15 00:00:00

First End Point

Sample Volume

10.0mL

End Point:

7.00pH


Time Delay:

10s

Concentration

of Titrant

End Point Switch:

ON

0.1000

PgDn

PgUp

Select Mode 01/02

01 HC1 ---﹥

NaOH(0.1)

02 K

2

Cr

2

0

7

---﹥

Fe2+(0.1)

Save as Delete Titrate Stir Calibrate

















Diag. 48 Display for Viewing Mode
On the left side is total number of mode and current selected mode, in mentioned diag.48, there are
total two modes and currently user selects No. 1 mode. On the bottom, there are ‘PgUp’ and ‘PgDn’,
user could turn over the page to display the mode.
On the right side is the relative parameter corresponding to selected mode, including Name of

Mode, mV/pH Titration, Number of End Point, Storing Time, End Point Value corresponding

to Each End Point, Time Delay, End Point Switch and Sample Volume, Concentration of

Titrant.
In current page, user could select mode directly, otherwise user could press ‘PgUp’ and ‘PgDn’,
button to display the other modes.
Please refer to chapter 7.3 Mode of Titration for the meaning of each mode parameter. User could
select and amend the parameter depends on actual need.

Note: User could not amend the system mode parameter provided by meter.

In viewing status, user is allowed to save as (copy), delete the mode, of course user is also allowed
to make mode titration. So in viewing mode, user is allowed to set stirring speed, calibrate electrode
slope.