7.4.4 End Point Jump Value Parameter
The same as End Point Jump Value Parameter in Pre-Titration Mode.
 
7.4.5 Pre-adding Volume Parameter 
The same as Pre-adding Volume Parameter in Pre-Titration Mode.
 
7.4.6 Each Adding Volume Parameter 
Blank titration mode is to add pre-adding volume first. When first adding finished, add the 
same volume each time in sequence. The adding volume every time is the adding volume 
after pre-adding volume. 
 
7.4.7 End Volume Parameter 
The same as Titration Mode End Volume Parameter 
When all above setting finished, press ‘Start Titration’ button to start blank titration. 
 
7.5 Manual Titration Mode 
It can be used to add titrant manually and to make titration manually. This titration mode will 
help user to find titration end point. In starting status, press ‘Titration’ button, select ‘Manual 
Titration’ to enter into manual titration mode. Displayed as follows: 
Manual Titration/Parameter setting 
 
 
 
 
 
 
 
 
 
 
 
Start Titration
 
MV Titration pH Titration 
Pre-adding: 10mL
Next Adding: 0.2mL
 
First end point 
Jump Value: Large
Middle
Small
Value: 100
End volume: 40mL
MV/pH Titration
Jump Value Set
Stirring
Resume Set
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 Diag.36 Parameter Setting in Manual Titration Mode 
 
All the same, Manual Titration Mode has following parameter: mV/pH Titration Parameter, 
Pre-adding Volume Parameter , Next Adding Volume Parameter ,End Point Jump Parameter,
End Volume Parameter. 
 
7.5.1 mV Titration Parameter 
The same as mV/pH Titration Parameter in Pre-Titration Mode
 
7.5.2 Pre-adding Volume Parameter 
The same as Pre-adding Volume Parameter in Blank Titration Mode.