LABEC M-TIT-5 User Manual

Page 46

Advertising
background image













Diag.62 Two-point Calibration in starting status

View Titrate Mode Set Calibrate

Calibration/Two-point calibration

9.18

pH


Titrate

25.0


Clean

Enter

Exit

Feed


Titration Panel READY + K 10/21

b. When displayed pH value stable, press ‘Enter’ button, meter displays ‘Calibration finished!’ and

calibrated electrode slope value, this means two-point calibration finished. Now calibration also
finished. Press ‘Enter’ button to quit from calibration panel. In the process of calibration, user
could stop calibration at any time by pressing ‘Exit’ button


12. pH Measurement

In the starting state, if the meter displays the potential value and temperature value, user can
press potential and temperature displaying area, the meter will turn to pH measuring status.
Press again, meter will return to potential displaying status. Displayed as diag.63













Diag. 63 PH displaying in starting status

View Titrate Mode Set Calibrate

1.00

pH

Titrate

25.0

Clean


13. Calibration of Burette Coefficient

In the starting state, clean burette for several times, fill the burette with distilled water (there
can't be bubbles in the burette). Set the burette and set the burette coefficient as 100%. Take a
clean weighing bottle, first weigh the empty bottle with the balance of 1/10000, put dripping
tube into the weighing bottle, press "Titration" button to enter into titration mode, move cursor
onto "Manual Titration", then press "Enter" to enter into manual titration mode. Please select

Titration Panel READY + K 10/21

Feed

45