V. maintenance and service of the meter – LABEC M-TIT-5 User Manual

Page 47

Advertising
background image

mV manual titration when titrate the parameter by manual, if the burette is 10ml, set pre-added
volume to "10ml"; if the burette is 20 ml, set pre-added volume to "20 ml" (ensures that end
volume is larger than pre-added volume), refer to diag.64















Diag.64 Parameter setting in Manual Titration

Manual Titration/Parameter setting

When setting finished, press ‘Start Titration’ button, meter will push the burette full of distilled
water into the weighing bottle, then weigh it with balance. Calculate the burette coefficient
according to following formula:

0

1

2

*V

d

g

g

f

×100%

In which: f= burette coefficient

g

1

= weight of empty weighing bottle

g

2

= weight of weighing bottle for adding solution

d= density of water

V

0

= volume of burette full scale

V. MAINTENANCE AND SERVICE OF THE METER

1. Maintenance
 Keep the sockets of the meter clean and dry, keep them away from acidity, alkaline, salt

solution, and also keep them damp-proof to ensure insulation and high input impedance
performance of the meter. When the meter is not working, insert Q9 short circuit plug into the
socket of measuring electrode to prevent it from dust and steam. On the occasion that is with
higher humidity, dry the plug of the electrode with clean gauze first.

 Often clean the whole burette with distilled water, especially the titrant which produces

sediment or crystal (eg. AgNO

3

), clean the burette in time after it is used.

 When using perchloric acid Acetic acid as the titrant, keep the ambient temperature ≥16℃,

otherwise it will produce crystal, and damage the valve.

2.Settle the common trouble: See Table 1


MV Titration pH Titration

Pre-adding volume: 10mL

Next adding: 0.2mL

Jump Value Set

mV/pH Titration

First end point

Jump value: Large

Middle

Small

Value :

100

Ending volume:

40mL

Resume Set

Stir

Start Titration