c. When pH value stable, press ‘Enter’ Button, the meter displays percentage slope. Now one-point

calibration has finished. Displayed as follows:














Diag. 60 One-point calibration finished slope in starting status

View Titrate Mode Set Calibrate


Press ‘Enter’, meter will ask user if you want to make two-point calibration, displayed as follows:
press ‘Enter’ to make two-point calibration, press ‘Exit’ to stop calibration and meter returns to
starting status. User could stop calibration at anytime by press ‘Exit’ button.












Diag.61 Display of whether making two-point calibration

11.2 Two-Point Calibration

Two-point calibration ensures pH measurement precise, which means the electrode system is
calibrated with two standard solutions to obtain actual percentage slope and Eo value of pH
combination electrode. The operation procedure is as follows:

a. When one-point calibration finished, the meter will ask you if you want to make two-point

calibration, press ‘Enter’ to enter into two-point calibration working status. Take out the electrode,
clean it with distilled water then, put it in pH standard buffer solution C. Meter displays current
pH value and temperature value.









Calibration/one-point calibration

Ca

e

librate electrode slop


Titration Panel READY + K 10/21

Enter

Exit

Calibration finished!
Electrode slope: 100.0%

Titrate

Enter

Clean

Feed

View Titrate Mode Set Calibrate








Calibration/one-point calibration

Calibrate electrode slope

Titration Panel READY + K 10/21

Enter

Exit

Two-point calibration?

Titrate

Enter

Exit

Clean

Feed