CONTROL PANEL OVERVIEW

Figure3: Main Control Panel

Power Switch

The main power switch on the control panel controls all power to the workstation and must be in the
on ( I ) position before any of the BACTRONEZ

’s systems are

powered. The switch will illuminate

when on.

Chamber Gas Light

This light indicates that anaerobic mixed gas is being injected into the workspace chamber.

Main Temperature Control and Green Digital Display

This control is marked SET TEMPERATURE and consists of up and down arrow buttons for
inputting a set point temperature and switching to calibration mode. The green digital display shows

the incubator’s

temperature in its default display, accurate to within + 0.1

C, as well as the

temperature set point when the arrow keys are being used to make set point adjustments. In
calibration mode, the control and display are used to enter temperature offsets to match the

BACTRONEZ’

s displayed temperature to the temperature detected by an independent calibration

(reference) thermometer.