Disinfecting, Vacuum pump maintenance, User maintenance (continued) – Shellab BACTRONEZ User Manual

Page 54

Advertising
background image

54 |

P a g e

USER MAINTENANCE (CONTINUED)

D

ISINFECTING

Disinfect the BACTRONEZ on a regular basis. Perform the following steps to disinfect the
workstation:

1. Turn the unit off. Open all doors and carryout your laboratory or production space

disinfection protocol.

2. After completion of your institutional protocol, allow all disinfectants to evaporate

completely. Wipe down all surfaces, except the acrylic glass panels, with distilled water and
Zephiran until the unit no longer has a volatile odor. Use nonabrasive wipes.

3. Remove all of the interior parts (shelves, racks, and any additional items) from the

incubators. Disinfect the workstation, including all corners, the incubator interiors, and the
pass box, using a suitable disinfectant. Take special care when cleaning around the pass
box door and arm port door gaskets so as not to impair the positive seal.

4. Disinfect the BACTRONEZ using commercially available disinfectants that are non-

corrosive, non-abrasive, and suitable for use on stainless steel surfaces. Contact your local
Site Safety Officer for detailed information on the proper disinfectants suitable for your
applications.

5. Do not use overtly volatile disinfecting agents. Chlorines, amphyls, and quaternary

ammonias will evaporate into the chamber environment. Over time the concentration in the
chamber atmosphere will continue to increase, potentially leading to inhibited growth,
reproduction, or metabolic symptoms in sample populations.

V

ACUUM

P

UMP

M

AINTENANCE

Refer to the operation manual supplied with your vacuum pump for recommended maintenance
routines.