Troubleshooting the presence of oxygen, Exiting the chamber, Exiting the chamber procedure – Shellab BACTRONEZ User Manual

Page 37: Procedure, Operation (continued)

Advertising
background image

37 |

P a g e

OPERATION (CONTINUED)

T

ROUBLESHOOTING

T

HE

P

RESENCE OF

O

XYGEN

1. Verify that proper sleeve donning, entry, and exiting procedures have been used.

2. Check if closed containers are being introduced through the pass box. Closed containers

can hold significant volumes of oxygen.

3. Seek to reduce amount of daily pass box usage.

4. Make sure that the arm port doors and sleeve assemblies are correctly installed.

5. Verify that the manometer is filled to the fill line with water.

6. Verify that gas input lines are attached to the BACTRONEZ

s GAS IN 1 and GAS IN 2

ports.

7. Make sure the foot pedal assembly center GAS / VAC line is securely connected to the

brass barb adaptors on both arm ports.

8. Check that the foot pedal assembly GAS and VAC lines are connected to the GAS OUT

and VAC OUT ports on the BACTRONEZ.

9. Turn the AMG cylinder regulator to 20 psi.

10. Uninstall the catalyst cartridge. If the cartridge is not warm, it may not be active, or it has

not been exposed to AMG and oxygen. Replace it with a second catalyst cartridge (Part
Number 9990759). Reactivate the first cartridge. Please see the

Reactivating the Catalyst

procedure on page 42.

Note: Use caution when uninstalling the catalyst cartridge. The catalyst may have grown hot during

the purge procedure.

11. Make sure that the anaerobic indicator strip packets have not been opened prior to being

placed in the oxygen-free atmosphere of the chamber.

Advertising