Gas conservation methods, Operation (continued) – Shellab BACTRONEZ User Manual

Page 44

Advertising
background image

44 |

P a g e

OPERATION (CONTINUED)

8. Allow the incubator a minimum of four (4) hour to stabilize after reaching the new

temperature.

9.

Compare the reference device’s reading with the incubator’s. If the reference device’s and
the incubator’s temperature readings are the same or fall within the range of your

laboratory protocol, the incubator is now calibrated.

10. If the two readings still fall outside your laboratory protocol, repeat steps 5

8.

11. If the temperature readings of the incubator and the reference device fall outside your

laboratory protocol after three calibration attempts, contact

Sheldon Technical Support

for assistance.

G

AS

C

ONSERVATION

M

ETHODS

1. Minimize the number of times per day the pass box is cycled.

2. Introduce small individual items such as, sealed micro plates or transport tubes, into the

workspace chamber through the sleeve assemblies.

3. Moving a large number of items through the pass box in one transport will reduce the

volume of AMG used in an auto or manual cycle.

4. When transporting a small number of items through the pass box, placing a large solid

object in the pass box can also help to reduce the volume of gas utilized.

5. Employ proper sleeve techniques when entering and exiting the workspace chamber.

6. Avoid making fast or large movements while working in the chamber. Use a swimming

motion, withdrawing one arm partly into the arm port while reaching in with the other.