Connecting to an in-house vacuum system, Connecting the vacuum pump, Filling the manometer – Shellab BACTRONEZ User Manual

Page 29: Installing the arm port doors, Connect the vacuum pump, E 29, Operation (continued)

Advertising
background image

29 |

P a g e

OPERATION (CONTINUED)

C

ONNECTING

T

O AN

I

N

-H

OUSE

V

ACUUM

S

YSTEM

Carry out this procedure to connect the BACTRONEZ to an in-house vacuum system. An in-house
system must be capable of evacuating a minimum of 25 inches (63.5 cm) of mercury.

1. Connect a 3/8 inch hose from the in-house system to the VAC IN port on the back, right

side of the BACTRONEZ.

C

ONNECTING THE

V

ACUUM

P

UMP

Use this procedure to connect a vacuum pump to the BACTRONEZ.

1. Connect the 3/8 inch hose from the vacuum pump to the BACTRONEZ

’s VAC

IN port.

Note: Plug the vacuum pump into the circular grey power outlet located on the back of

the BACTRONEZ at the top right. Do not plug into a wall outlet! Plugging the
pump into a wall outlet will run the pump non-stop, rather than only when needed
by the BACTRONEZ.

2. Ensure that the vacuum pump is placed in a ventilated location to prevent

overheating. An overheated vacuum pump can become a fire hazard.

3. Make sure there are no kinks in the vacuum line running from the pump to the

BACTRONEZ.

F

ILLING THE

M

ANOMETER

1. Fill the manometer with distilled water to the refill ring (the top of the two

reference rings). This can be done by unscrewing and removing the
manometer bottle, or pouring water in through the hole on the top of the
manometer assembly.

2. Distilled water is preferable to tap water to avoid scaling deposits from

waterborne minerals.

I

NSTALLING THE

A

RM

P

ORT

D

OORS

The arm port doors should already be inside the workspace chamber.

1. Turn the silver locking bar to a roughly 45° position.

2. Install each door bottom first, inserting the tabs into the slots on the arm ports.

3. Tilt the door cover up so that it fits securely within the arm port.

4. Turn the locking bar to the horizontal position:

5. Secure the door by turning the black arm port door knob clockwise, using wrist

strength only, until the knob grabs and feels snug. Tightening too much may
compromise the integrity of the door by pulling the post that the knob and
locking bar are mounted on out of position.

Figure 17: Vacuum

Pump

Figure 18: Manometer

PN 9900699 / PN 9900698

Figure 19: Arm Port

Door