OPERATION (CONTINUED)

E

STABLISHING AN

A

NAEROBIC ENVIRONMENT

Carry out the following steps to purge the aerobic atmosphere from the workstation and establish
an anaerobic environment. If the BACTRONEZ has previously been in use, remove the right arm
port sleeve prior to starting the procedure.

This procedure requires approximately 1 to 4 hours to complete.

1. Verify the following:

a. The interior and exterior pass box doors are closed and secured.

b. The manometer is filled to the fill line (upper ring) with water.

c. The foot pedal assembly center line is securely connected to the brass barb

adaptors on both arm port door assemblies.

d. The foot pedal assembly GAS and VAC lines are connected to the GAS OUT

and VAC OUT ports on the BACTRONEZ.

e. Gas input lines are attached to the BACTRON

EZ’

s GAS IN 1 and GAS IN 2

ports.

2. Note the reading on the AMG gas regulator.

a. The BACTRONEZ workstation will require approximately 200 psi of AMG for a

setup purge.

3. Make sure the AMG cylinder regulator is set to 15 to 20 psi. Open the valve all the way to

start a flow of gas to the BACTRON

a. Nitrogen and other inert gasses should not

be used to purge the workstation.

4. Turn on the workstation, if it is not already powered up.

5. Open the right arm port door.

6. Verify that all incubator doors are slightly open to avoid

leaving reservoirs of aerobic atmosphere.

7. Unwrap and install the catalyst cartridge on the left wall

inside the workspace chamber. See Figure 22 to the right.

a. Catalyst cartridges come ready for use from

the factory. However, if you have stored the
cartridge for longer than six months, it is
strongly recommended that you reactivate it prior to
its first use. Please see

Reactivating the Catalyst

Cartridge section

on page 42.

b. The catalyst cartridge will grow warm in the presence of oxygen and AMG.

Procedure continued on next page

Figure 22: Catalyst Cartridge
Installed