CONTROL PANEL OVERVIEW (CONTINUED)

Heating Activated Light

The clear pilot light located beneath the label HEATING ACTIVATED illuminates whenever the
heating elements are powered and warming the incubator.

Set Over Temperature

This graduated dial sets the

incubator’s

temperature limit for the Over Temperature Limit backup

system. The OTL System operates independently of the Main Temperature Control, and prevents
uncontrolled heating of the incubator in the event of a failure of the Main Controller while in its
heating mode. For more details, please see the explanation of the

Over Temperature Limit

System

in the Operation section

’s Theory of Operation.

OTL Light

This pilot light is marked OVERTEMPERATURE ACTIVATED and is directly below the Over
Temperature Limit control. The light will turn on when the OTL System has taken control of the
incubator, and will be accompanied by a buzzer alarm. Under normal operating conditions this light
should not illuminate.

Pass Box Doors Light

The clear Doors pilot light indicates that both pass box doors are closed. This light must be on in
order to cycle the pass box.

Anaerobic Light

This is light activates after the completion of a full auto cycle of the pass box Interior.

Aerobic Light

This light activates when the outer pass box door has been opened, exposing the pass box to
aerobic atmosphere. It remains on until the next completion of a full auto cycle.