Control panel overview (continued) – Shellab BACTRONEZ User Manual

Page 20

Advertising
background image

20 |

P a g e

CONTROL PANEL OVERVIEW (CONTINUED)

Pass Box Vacuum Controller and Digital Display

The main control panel vacuum gauge-controller displays the level of vacuum in inches of mercury
in the pass box interior. It controls the application of vacuum to the pass box during the pass box
auto cycle.

Figure 4: Pass Box Vacuum

Controller and Display

Automatic Start / Reset Switch

The all black Automatic Start / Reset switch initiates or aborts the pass box’s auto vacuum

-and-

gas-recharge cycle used to remove aerobic atmosphere from the pass box. Pressing then releasing
on the left (Start) the will initiate the auto cycle. Pressing and holding on the right (Reset) for three
seconds will abort and then restore the pass box to a near standard atmospheric pressure sufficient
that the pass box doors can be easily opened.

Pass Box Gas Light

This light indicates that the anaerobic mixed gas is being injected into the pass box during the auto
cycle. The pass box Gas light does not illuminate during manually gas injections.