27 |

P a g e

OPERATION (CONTINUED)

Figure 12: Gas and Vacuum Ports

S

ET UP THE

G

AS

S

UPPLY

See

the Gas Source entry

in the Installation section on page 15 for Sheldon

Manufacturing’s recommended gas applications.

1. Install the gas regulator or regulators on the compressed gas cylinders

you will be using.

2. The regulators should be set to 15 to 20 psi. Please see

the Pressure

Unit Conversion table

in the Operation Section for working with units

other than Pounds per Square Inch (page 46).

Warning

: Never exceed a 5% hydrogen concentration inside

the unit’

s chamber.

Avertissement:

La concentration d’

hydrogène ne doit pas dépasser

5%

dans la chambre anaérobie.

Figure 13: Hose

Connections

Figure 14: Hydrogen
Warning Label

To Vacuum Pump or

In-House Source

To Foot Pedal VAC

Port

To Foot Pedal

GAS Port

To AMG Cylinder

(Single Gas

Application)