OPERATION (CONTINUED)

Accessing the Workstation

Items such as media containers and laboratory equipment can be introduced to or removed from
the BACTRONEZ

’s

anaerobic environment through the pass box. The pass box creates a near

anaerobic environment through three cyclic applications of the BACTRONEZ

’s vacuum pump

to

reduce air volume in the pass box, followed by gas replenishment injections. The pass box
operates in one of two modes: A user-initiated three stage auto-cycle, or through the use of the
manual control switch to lower and then recharge the atmosphere of the pass box space with
anaerobic mixed gas three times. The option to manually cycle the pass box is primarily intended
as a backup for the auto cycle system.

Users can access and work glove-free in the workspace chamber by donning the sleeve
assemblies attached to the front panel arm ports. The sleeve assemblies are also compatible with
exam gloves for handling pathogenic samples inside the workspace. After being donned

but prior

to opening the arm port doors

the sleeve assemblies are cycled by using the foot pedal assembly

to apply reduced pressure to the sleeves with the vacuum pump, then to charge the sleeves with
AMG.

Purging, sealing, and effective use of the sleeve assemblies requires bare skin contact between the

widest part of the user’s forearms and the

cuff ring of the sleeve assembly. Smooth, small items

held in hand may be introduced into the workspace chamber through the sleeve assemblies.

Incubators

The BACTRONEZ is provided with a cabinet style incubator in the workspace chamber. The
incubator

’s temperature is controlled by a

digital controller board using a solid state temperature

sensor probe attached to the incubator body and a heating element. When heating the incubator,
the controller uses Proportional

Integral

Derivative (PID) analytic functions to slow the rate of

heating as it approaches the user-selected set point. This is done to avoid overshooting, and to
correct for naturally occurring measurement errors.

The digital controller also uses PID learning functions to optimize its warming rates for hotter or
cooler environments. If the BACTRONEZ is moved to a new location with a significant temperature
difference from its previous surroundings, it may require 24 hours of incubator run time for the
controller to fully adapt to the new thermal environment and obtain its previous high level of
temperature stability. This is why the incubator should be run at its application set point for 24 hours
prior to calibration. Additionally, the heat loss from leaving the incubator doors open for long
periods of time (an hour or more) can trick the controller into thinking it is operating in a cool
environment. This can result in a period of temperature overshooting.