Directions to calibrate do sensor %sat value – YSI 5200A User Manual

Page 40

Advertising
background image

YSI 5200A

79

YSI 5200A

78

9. When the conductivity and temperature values show no significant change

for approximately 30 seconds, press

to continue calibrating OR

press

to abort the calibration and return to the calibration menu

without saving any changes.

10.

Use the arrow keys to select numbers on the numeric keypad, enter

the value of the calibration solution. Press

to confirm each number

selection.

Note:

- The conductivity value of the calibration solution varies with temperature.

For proper calibration, enter the value of the calibration solution as it is

listed for 25°C. The 5200A is programmed to calibrate to this temperature

compensated conductivity value which is commonly referred to as Specific

Conductance.

- The calibration value range is automatically determined when the sensor

is placed in calibration solution. The four ranges are: 0-200µS, 0-2000 µS,

0-20mS, and 0-200mS.

- Take note of the calibration solution’s unit of measure and compare it to the

unit of measure selected by the 5200A’s auto range function. If the instru-

ment’s selected range is in mS, then the value of the calibration solution must

Configuring the 5200A

be entered as mS, not µS. 1,000 µS = 1 mS. For example, if the solution is

10,000 µS at 25°C and the 5200A has selected 0-20 mS for the calibration

range, enter 10 mS as the calibration solution value.

11. Press

to confirm the entered value OR press

to abort the

calibration and return to calibration menu without saving any changes.

12. Press

to save the calibration and return to the Run Screen.

13. Remove the calibration/storage cup from probe assembly.

14. Rinse sensor and calibration/storage cup with clean water before calibrat-

ing the next sensor system and before reinstalling the probe assembly in

the system water.

DO

Menus → Calibration→ DO

To determine which sensor is the DO sensor, See YSI probe assembly 5562 - figure

4.5, page 76 or YSI probe assembly 5561 - figure 3.11, page 35.

Calibrate the DO sensor using a one point calibration. Calibration can be done in

% Saturation (%Sat) or mg/L. Calibrating in %Sat automatically calibrates mg/L

and vice versa so it is only necessary to calibrate one or the other, not both. For

both ease and accuracy, YSI recommends calibrating in %Sat.

Salinity compensation for mg/L readings is obtained from either the conductivity

sensor, if installed and enabled - page 153, or from the user entered ppt value in

the DO sensor setup menu - page 147.

Directions to Calibrate DO Sensor %Sat Value

1. Connect Probe Assembly - page 61.

2. Read Getting Ready to Calibrate - page 75.

3. Wait at least 10-15 minutes after the 5200A is powered on, or after the

DO sensor is serviced, before calibrating. (This allows enough time for

the DO sensor to stabilize).

4. If not using a Conductivity sensor or if the conductivity sensor is disabled,

enter the salinity (ppt) value of the water to be monitored in DO Sensor

Setup menu. This will result in accurate DO mg/L readings.

Menus → Sensor Setup→ DO→ppt

5. Enter the local elevation in the DO Sensor Setup menu, if not already

configured.

Menus → Sensor Setup→ DO→Elevation

Configuring the 5200A