Clear logs, Date/time – YSI 5200A User Manual

Page 55

Advertising
background image

YSI 5200A

109

YSI 5200A

108

3.

Press

to select.

4. Press

to turn off or on.

5. Press

to save configuration and return to Run Screen.

Press

to view the 25 most recent event logs.

Press

to view active (“on”) non timer events. See Control Menu - page 166

for additional information.

Notes:

- The General Alarm does not create an event log whether a general alarm

relay is enabled or not.

- Events are not recorded when relays de-energize due to systemwide events.

Systemwide events include: calibration, Svr Hold, factory resets, power

cycle, 30 second sensor system reset, relay test, and disabling systems.

- Event logs record the default system name and not the user configured

system name.

- Event logs do not record relay number or user configured relay name.

- Aux digital events will never display “alarm” regardless if alarms are

enabled, see below screen shot on right.

Clear Logs

Menus → System → Data Logging → Clear Logs

Use the Clear Log option to permanently delete all event and/or sensor log records

from the 5200A.

Before permanently deleting sensor and event logs, poll all data using AquaManager.

AquaManager will store all logs to a database in PC database.

Configuring the 5200A

1. Enter Clear Log menu

Menus → System → Data Logging → Clear Logs

2.

Use

and

to scroll submenu.

3.

Press

to select.

4. The instrument will prompt, “Are You Sure?” Press one of the following

softkeys to confirm:

.

5. Selecting “Yes” deletes logs after approximately three seconds.

6. Selecting “No” or “Cancel” does not delete logs and returns to the Data

Logging menu.

7. Press

to return to Run Screen or

to System menu.

Notes:

- Logs cannot be deleted from the 5200A instrument using AquaManager.

- Calibration and Configuration

logs cannot be cleared from 5200A.

- Clear data logs are updated immediately to the system and not when

is pressed. YSI recommends pressing

in this instance

in order to save any other changes that may have been made to the con-

figuration.

Date/Time

Menus → System → Date/Time

It is important to configure the correct date and time in the 5200A. Four 10-event

daily timers and one feed timer run based on the 5200A date and time. Logs record

events based on the 5200A clock. Date and time display formats are user defined.

Configuring the 5200A

Advertising