Directions to calibrate orp sensor, Directions to calibrate ph sensor – YSI 5200A User Manual

Page 42

Advertising
background image

YSI 5200A

83

YSI 5200A

82

11. Press

to save the calibration and return to the Run Screen.

12. Rinse sensor in tap or purified water and place probe assembly in system

stream or continue calibrating other sensors.

ORP

Menus → Calibration→ ORP

The ORP sensor - figure 4.5, page 76 - is calibrated with a one point calibration.

Directions to Calibrate ORP Sensor

1. Connect probe assembly - page 61

2. Read Getting Ready to Calibrate - page 75

3. Fill calibration cup with 30 ml of known ORP solution. We recommend

using Zobell solution.

4. Carefully immerse the sensor into the solution.

5. Gently rotate and/or move the probe module up and down to remove any

bubbles from the pH/ORP sensor.

6. Screw the calibration/storage cup on the threaded end of the probe

module.

7. Enter ORP calibration menu:

Menus → Calibration→ ORP

8. Allow at least one minute for temperature equilibration.

9. After one minute, verify the ORP and temperature values show no sig-

nificant change for approximately 30 seconds. Then, press

to

complete calibration OR press

to abort the calibration and return

to calibration menu without saving any changes.

10.

Use the arrow keys to select numbers on the numeric keypad, enter

the value of the solution. Press

to confirm each number selection. The

Configuring the 5200A

ORP solution value varies with temperature. Note the current temperature

reading of the solution and enter the solution value that corresponds to

that temperature. The calibration solution should include a temperature

chart that contains solution values based on the temperature. Only whole

numbers can be entered.

11. Press

to confirm the entered value OR press

to abort the

calibration and return to calibration menu without saving any changes.

12. Press

to save calibration and return to Run Screen.

13. Remove the calibration storage cup from probe assembly.

14. Rinse the sensor in tap or purified water before continuing to calibrate

other sensors or reinstalling the probe guard.

pH

Menus → Calibration→ pH

The pH sensor - figure 4.5, page 76 - can be calibrated using either a one or

two-point calibration. For the highest accuracy, YSI recommends performing a

two-point calibration. Use two pH buffers that cover the pH range of the system

being monitored. For example, if the pH of the system water varies between 5.5

and 7, perform a 2-point calibration using pH 7 and pH 4.

Directions to Calibrate pH Sensor

1. Connect probe assembly - page 61.

2. Read Getting Ready to Calibrate - page 75.

3. Fill the calibration cup with 30 ml of pH buffer 7 solution.

4. Carefully immerse the sensor into the solution.

5. Gently rotate and/or move the probe module up and down to remove any

bubbles from the pH or pH/ORP sensor.

6. Screw the calibration/storage cup on to the threaded end of the probe

module.

7. Enter pH calibration menu:

Menus → Calibration→ pH→2 Point Cal (recommended)

OR

Menus → Calibration→ pH→1 Point Cal

8. Allow at least one minute for temperature equilibration.