Sensor care and maintenance, Do sensor – YSI 5200A User Manual

Page 93

Advertising
background image

YSI 5200A

185

YSI 5200A

184

CAUTION: The pH sensor should never be exposed to subfreezing air temperatures

for more than a few minutes.

5. Remove the probe guard from the probe.

6. Remove any debris from the sensors and then rinse carefully in a bucket

of clean water, being very careful not to damage the sensors.

7. 5562 probe assembly only: Using the small round brush provided with

the sensor, repeatedly scrub the two open holes in the conductivity sensor.

See figure 4.5 page 76.

8. While the probe guard is removed, inspect the dissolved oxygen membrane

for obvious holes or tears. Do not remove the membrane cap at this time.

9. Replace the probe guard.

10. Exit Service Hold if Service Hold timer has not expired. See pages 128

and 169 for Service Hold information .

11. 5562 probe assembly only: Place the probe in enough pH 7 buffer to

immerse the pH and temperature sensor. Wait about 3 minutes for the

readings to stabilize and record the reading shown on the 5200A. The

deviation from pH 7 will reflect the sensor drift since the last calibration.

12. 5562 probe assembly only: If necessary, i.e. pH sensor exhibits large drift

from last calibration, perform a two-point pH calibration. See pH calibra-

tion starting on page 83.

13. Rinse the probe with water and then place it into the calibration/storage

cup containing about 1/8 inch (3 mm) of water or a moisten sponge. Make

sure that the dissolved oxygen sensor is not immersed in the water and

that there are no water droplets on the membrane.

14. Wait about 5 minutes for temperature equilibration and then record the

dissolved oxygen reading in percent air saturation as shown on the 5200A

display.

15. If the dissolved oxygen readings show minimal calibration drift (i.e.,

within about 5% of the correct value) and the readings are stable, proceed

to Step #18. If excessive drift is noted or the readings are jumpy, replace

the membrane cap as described in the instruction sheet included in the

Membrane Kit.

16. After replacing the membrane, calibrate the DO sensor. See calibrate DO

sensor starting on page 79.

17. 5562 probe assembly only: Validate ORP value by placing probe assembly

in solution of known ORP value. If ORP sensor exhibits large drift from

last calibration, perform sensor calibration. See ORP calibration page 82.

18. Place the probe back in the sample stream and continue monitoring.

19. Test all control and alarm relays for proper operation.

NOTE: This procedure does not include quantitative data taken with regard to

either the temperature or conductivity sensors. The temperature sensor only fails

Maintenance

in very rare circumstances and, if it does fail, the temperature readings will seem

unreasonable for the current conditions. In the unlikely event that the temperature

sensor shows unusual and/or jumpy readings, it will need to be checked by a YSI

authorized service center. No user service is possible. For the conductivity sensor,

drift is usually minimal except as caused by build-up of debris in the sensor. Once

the sensor is cleaned as described in Step 5, the sensor almost always yields accurate

readings. However, if you wish to check your conductivity calibration, place the

probe in a known standard being sure that the top vent hole is completely immersed.

If significant drift has occurred, recalibrate the sensor.

Alternative Quality Assurance Protocol (quick check and

adjustment)

An alternative protocol, complementary to the quality assurance program already

described, can be carried out by comparing the current dissolved oxygen and pH

readings shown on the 5200A with those taken by a recently calibrated hand held

instrument. For dissolved oxygen, the probe of a handheld DO instrument can

be placed in the stream near the 5200A probe. If a comparison of the readings

from the recently calibrated DO instrument and the 5200A indicates calibration

drift, the 5200A can be recalibrated. Note - this comparison should only be used

to determine if a calibration is necessary once the 5200A probe has been cleaned

of all algae and other debris. If a gross variation is observed between the two

instruments (> 1 mg/L), it is likely that maintenance is required on the 5200A DO

sensor.

For pH, a similar field calibration adjustment can be made using a field pH meter.

Sensor Care and Maintenance

Periodic sensor cleaning and DO membrane changes are required.

DO Sensor

The DO sensor requires periodic membrane changes and electrode cleaning. For

best results, YSI recommends that the electrolyte (02 probe) solution and the

membrane cap be changed at least once every 30–90 days. Refer to the instruc-

tion sheet included with the membrane kit for specific instructions on changing a

membrane.

Notes

- Membrane life depends on usage. Membranes will last a long time if

installed properly and treated with care. Erratic readings are a result of

Maintenance

Advertising