1. Enter Date/Time menu
Menus → System → Date/Time
2. Verify 5200A current date and time.
3. If necessary, use
▲
and
▼
to scroll and highlight submenus.
4.
Press
to select.
5. Configure submenus.
6. Press
to save setting and return to Run Screen or press
to System menu.
Notes:
- Changes to Date and Time menus take place immediately and are saved
prior to
.
- Configured date and time formats are carried throughout menus and
logs.
- Any active (with energized relay) timer system, including 10-event daily
timers, feed timers, and/or dosing timers, are NOT recalculated when
the 5200A time and/or date is changed. Once a timer relay has been
energized, it runs for the duration and does not reset based on changes
to the 5200A date and time. Example - timer 1, event 1 is configured to
come on at 9:00am for 10 minutes. Timer relay energizes at 9:00am. At
9:05am the 5200A time is changed to 10:05am (for daylight savings ad-
justment). The timer relay will de-energize after five additional minutes.
The 5200A time will be 10:10am. The event log will reflect the configured
5200A time when the event occurred. In this example, 9:00am for the
start and 10:10am for the end.
- The time must be manually changed for daylight savings time adjustments.
Daylight savings note: Sensor data log records are not overwritten when
the clock is changed unless the instrument is power cycled after changing
the data and time. Example - the data log interval is set to 15 minutes.
Current instrument time is 6:00 am. The clock is adjusted to 5:00 am for
Configuring the 5200A
daylight savings time. The 5200A sensor log contains four 15 minute in-
terval records from 5:00 am to 6:00 am from prior to time change. Sensor
data logs are not overwritten with new values for the repeat of 5:00 am to
6:00 am 15 minute intervals unless the instrument is power cycled.
- Menu configurations are saved to the system by pressing
to return
to the Run Screen and not when the change is made in the menu. Excep-
tion - Date and time and clear data logs are updated immediately to the
system, not when
is pressed.
Display Settings
Menus → System → Display Settings
Sensor and Aux conditions are displayed at the Run Screen. Use the Display Set-
tings menu to configure how many systems display on one page at the Run Screen.
Up to six systems can be displayed on one Run Screen page. Multiple Run Screen
pages scroll at user configured speed (1-65535 seconds).
Configuring the 5200A