Serial number, Service, Reset clean probes – YSI 5200A User Manual

Page 62

Advertising
background image

YSI 5200A

123

YSI 5200A

122

Disabling the Power Fail System will disable AC power events and disable the General

Alarm in the event of low battery for DC power. Disable Power Fail AC Monitor

when powering 5200A by DC power source.

Enable Low Battery Monitor to trigger the General Alarm in the event DC voltage is

≤ to low battery trigger. The Low Battery Alarm Hold Off time is five (5) seconds and

cannot be user configured. Once the low battery general alarm has been acknowl-

edged by pressing one of 4 softkeys, it is not reset. See General Alarm - page 117.

1. Verify AC and DC power supplies are properly wired and powered - page

42. and page 129.

2. Enter Power Fail Alarm menu and submenus

Menus → System → Power Fail

3.

Use

and

to scroll and highlight submenus.

4.

Press

to select.

5. Enable and configure submenus.
6. Press

to save configuration and return to Run Screen.

Notes:

- Press

to view AC power fail events.

-

icon at the Run Screen indicates that the 5200A DC power is ≥ Low

Battery Trigger value with Low Battery Monitor is enabled.

-

icon at Run Screen indicates DC voltage ≤ the Low Battery Trigger

when Low Battery Monitor is enabled.

- Low battery alarm is not reset when front panel softkey is pressed.

- AC power fail does not generate a General Alarm.

Configuring the 5200A

Serial Number

Menus → System → Serial Number

The serial number is displayed at the serial number menu. The serial number is

required when contacting YSI customer service. The serial number is also neces-

sary to configure AquaManager. 5200A serial numbers begin with “53-”followed

by seven digits. Serial numbers are generated at the factory and cannot be changed

by the user.

Service

Menus → System → Service

The Service menu includes submenus to configure and Reset Clean Probe Timer,

Reset Systems to factory defaults, and test Relay(s).

Reset Clean Probes

Menus → System → Service→ Reset Clean Probes

Enable the clean probe timer, configure the duration, and reset clean probe days

remaining from Reset Clean Probe menu. Selecting reset at this menu resets the

clean probe timer to the user configured value.

icon displays at Run Screen

when the clean probe timer expires.

Configuring the 5200A