AquaManager and the System Menu

AquaManager users can configure most 5200A system menus from a PC. After poll-

ing the 5200A, use “Systems Management” from the node window to view and upload

5200A configurations. System menus are configured at the Setup and Input tabs.

Configuring the 5200A

Notes:

- Always manually poll the 5200A after uploading configuration changes

and verify the uploaded information is retained. When configuration

changes are uploaded using AquaManager at the same time configura-

tion changes are being made by a user in the menus on the 5200A, the

AquaManager uploads may not be saved depending on the timing of

keystrokes at the 5200A.

- The Run Screen must be displayed during normal operation. If the

5200A is not at the run screen during AquaManager uploading, pressing

at the 5200A will only save certain configuration changes.

See Chapter 5 AquaManager - page 172 and AquaManager online help for ad-

ditional information.

Sensor Setup

Menus → Sensor Setup→Aux 1

→Aux

2

→Conductivity

→DO

→ORP

→pH

→Temperature

→Salinity

Aux, Conductivity, DO, ORP, pH, and temperature system set points, control mode

(set point or PID/PWM), alarm and sensor system relays are configured in the

Sensor Setup menu. In addition, the salinity system is enabled in this menu. Four

onboard relays can be configured for control and alarm output devices. Enabled

relays energize when a control or alarm system is active. Alarm notifications can

be sent via email or cell phone (SMS messaging) - page 95.

Configuring the 5200A