Relay test – YSI 5200A User Manual

Page 64

Advertising
background image

YSI 5200A

127

YSI 5200A

126

1.

Use

and

to scroll and highlight submenus.

2.

Press

to select.

3. When prompted, “Are you sure?”, select one of the following to confirm,

4. Press

to save configuration and return to Run Screen.

Reset Sensors

Menus → System → Service→ Reset To Factory→ Reset Sensors

Specific sensors, aux system, or all sensor and aux systems listed in the Reset Sensors

menu can be reset to factory default.

1.

Use

and

to scroll and highlight submenus.

2.

Press

to select.

3. When prompted, “Are you sure?”, select one of the following to confirm,

4. Press

to save configuration and return to Run Screen.

Note:

- Perform a sensor calibration after resetting a sensor to factory default.

Configuring the 5200A

Reset Timers

Menus → System → Service→ Reset To Factory→ Reset Timers

A specific timer, feed timer, or all timers listed in the Reset Timers menu can be

reset to factory defaults.

1.

Use

and

to scroll and highlight submenus.

2.

Press

to select.

3. When prompted, “Are you sure?”, select one of the following to confirm,

4. Press

to save configuration and return to Run Screen.

Relay Test

Menus → System → Service→ Relay Test

Use the Relay Test menu to verify peripheral devices are wired correctly. Entering

this menu turns off any active relays and disables any sensor, aux, and timer system

relays from energizing regardless of value or condition. Systems are not reset when

the Relay Test menu is exited. To reset systems after exiting the Relay Test menu,

power cycle the 5200A. Example - an energized conductivity low control relay (set

point control normal mode) is de-energized when entering the Relay Test menu. If

the conductivity value is in control low range when the Relay Test menu is exited,

the relay will not re-energize until the system has a change of condition. The con-

ductivity system must reach set point, then fall in low control range again before

energizing the control low relay. It is strongly recommended to turn the power off,

then on, at the 5200A after exiting the Relay Test menu.