Operations

21

PXIe-9848

serves as a master module and can output trigger signals to the
PXI Trigger Bus Numbers 0 through 7.

Analog Trigger

PXIe-9848 analog trigger circuitry can be configured to monitor
one analog input channel from which data is acquired. Selection of
an analog input channel as the analog trigger channel does not
influence input channel acquisition. The analog trigger circuit gen-
erates an internal digital trigger signal based on the condition
between the analog signal and the defined trigger level.

Analog trigger conditions are either positive-slope trigger, in which

the trigger event occurs when the analog input signal changes
from a voltage lower than the specified trigger level to a voltage
exceeding the specified trigger level, or negative-slope trigger, in
which the trigger event occurs when the analog input signal
changes from a voltage exceeding the specified trigger level to a
voltage lower than the specified trigger level.

Figure 3-6: Analog Trigger Conditions

The trigger signal can be chosen from among CH0, CH1, CH2,

CH3, CH4, CH5, CH6 and CH7 while using an external analog

Trigger Level

Positive-Slope Trigger Event

Occurs

Negative-Slope Trigger

Event Occurs

Analog

Signal