Figure 3-13: Basic Digitizer Acquisition Timing

To achieve sampling rates other than 100MS/s, a number for scan

interval counter needs only be specified. For example, if the scan
interval counter is set as 2, the equivalent sampling rate is
100MS/s / 2 = 50MS/s. If as 3, the equivalent sampling rate is
100MS/s / 3 = 33.33MS/s, and vice versa. The scan interval coun-
ter is 16 bits in width, therefore the lowest sampling rate is
1.025KS/s (100MS/s / 65535).

Figure 3-14: Varying Sampling Rates by Adjusting Scan Interval Counter