1 post trigger mode, 2 pre-trigger mode, Post trigger mode – ADLINK PXIe-9848 User Manual

Page 33: Pre-trigger mode, Figure 3-7, Post-trigger acquisition, Figure 3-8, Pre-trigger acquisition

Advertising
background image

Operations

23

PXIe-9848

3.4.1

Post Trigger Mode

Post-trigger acquisition is applicable when data is to be collected
after the trigger event, as shown. When the operation starts,
PXIe-9848 waits for a trigger event. Once the trigger signal is
received, acquisition begins. Data is generated from ADC and
transferred to system memory continuously. The acquisition stops
once the total data amount reaches a predefined value.

Figure 3-7: Post-Trigger Acquisition

3.4.2

Pre-trigger Mode

Collects data before the trigger event, starting once specified func-
tion calls are executed to begin the pre-trigger operation, and
stopping when the trigger event occurs.If the trigger event occurs
after the specified amount of data has been acquired, the system
stores only data preceding the trigger event by a specified
amount, as follows.

Figure 3-8: Pre-trigger Acquisition

Time

Operation

start

Trigger

N samples

Data

Trigger Event Occurs
Acquisition start

Acquisition stop
Begin to transfer data to system

Time

Operation start
Acquisition start

Trigger

Data

Trigger Event occurs
Acquisition stop
Data transfer to system begins

X samples

X samples acquired
Despite acquired samples not reaching
N, acquisition stop and X samples in
memory are returned to system.

Specified amount of data,
(N samples)

Advertising