Table 3-5: Counter Parameters and Description
Counter 
Name
Length
Valid Value
Description
ScanIntrv
16-bit
1 - 65535
Timebase divider to achieve
equivalent sampling rate of the 
digitizer, where Sampling rate = 
Timebase / ScanIntrv
DataCnt
32-bit
1 - 536870911 Specifies the amount of data to
be acquired
trigDelayTicks 16-bit
1 -65535
Indicates time between a trigger 
event and commencement of 
acquisition. The unit of a delay 
count is the period of the 
Timebase.
ReTrgCnt
32-bit
1 -536870911 Enables re-trigger to accept
multiple triggers. Please see 
Section 3.4.5: Post-Trigger or 
Delayed-Trigger Acquisition with 
Re-Triggering for more details.