Table 3-5: Counter Parameters and Description

Counter
Name

Length

Valid Value

Description

ScanIntrv

16-bit

1 - 65535

Timebase divider to achieve

equivalent sampling rate of the
digitizer, where Sampling rate =
Timebase / ScanIntrv

DataCnt

32-bit

1 - 536870911 Specifies the amount of data to

be acquired

trigDelayTicks 16-bit

1 -65535

Indicates time between a trigger
event and commencement of
acquisition. The unit of a delay
count is the period of the
Timebase.

ReTrgCnt

32-bit

1 -536870911 Enables re-trigger to accept

multiple triggers. Please see
Section 3.4.5: Post-Trigger or
Delayed-Trigger Acquisition with
Re-Triggering for more details.

Advertising