10
Introduction
1.3.4
General Specifications
Bus and Physical
Bus interface
PCI, 32 bit/ 33MHz
PCI Bus Signaling
Universal PCI, support 3.3 V and 5 V PCI
signals
Physical dimensions
160 W x 100 H mm (6.3 x 3.94 in.)
Environment Requirements
Operating environment
Temperature:0°C - 50°C
Relative humidity: 5% - 95%, non-condensing
Storage Environment
Temperature: -20°C - +80°C
Relative humidity: 5% - 95%, non-condensing
Calibration
Onboard reference
+5 V
Temperature coefficient
≤ ±5 ppm/°C
Self-calibration
On software command, the PXI-9527 corrects
offset and gain error relative to internal high
stability, high precision reference
External calibration
interval
1 year
Warm-up time
15 minutes
Power Consumption
Power Rail
Standby Current (mA)
Full Load (mA)
+5 V
930
2330
+12 V
310
350