background image

NuDAM-6521

75

@Example

User command:

~300<CR>

Response:

!3000$#%@~*<CR>

Command leading code setting is $#%@~* for module address ID 30, current
status is factory default setting.

5.4.19 Change Command Leading Code Setting

@Description
User can use this command to change command leading code setting as he
desired.

@Syntax
~(Addr)10(C1)(C2)(C3)(C4)(C5)(C6)<CR>

~

Command leading code.

(Addr)

Address ID, range (00 - FF).

10

Change command leading code setting.

(C1)

Leading code 1, for read configuration status, firmware version,
etc. default is $. (1-character)

(C2)

Leading code 2, for read synchronize sampling, digital
output ,default is #. (1-character)

(C3)

Leading code 3, for change configuration.
default is %. (1-character)

(C4)

Leading code 4, for read alarm status, enable alarm, etc. default is
@. (1-character)

(C5)

Leading code 5, for read command leading code, change leading
code, etc. default is ~.
(1-character)

(C6)

Leading code 6, this leading code is reserved. default is *.
(1-character)

@Response
!(Addr)< CR>

or

?(Addr)<CR>

!

Command is valid.

?

Command is invalid.

(Addr) Address

ID.