SCPI Command Reference
High Speed Power Meter
163
Product-Specific Commands—Description
:SENSe[1..n]:CORRection:FACTor
[:MAGNitude]
Description
This command sets a correction factor. The units
are w/w by dÈfault.
Syntax
:SENSe[1..n]:CORRection:FACTor[:MAGNitude]<
wsp><CorrectionFactor[<wsp>W/W|DB]>|M
AXimum|MINimum|DEFault
Parameter(s)
CorrectionFactor:
The program data syntax for <CorrectionFactor>
is defined as a <numeric_value> element
followed by an optional <SUFFIX PROGRAM
DATA> element. The allowed <SUFFIX
PROGRAM DATA> elements are: W/W|DB. The
<CorrectionFactor> special forms MINimum,
MAXimum and DEFault are accepted on input.
MINimum allows to set the instrument to the
smallest supported value.
MAXimum allows to set the instrument to the
greatest supported value.
DEFault allows the instrument to select a value
for the <CorrectionFactor> parameter.
The <numeric _value> parameter is a
correction factor using W/W as units. If an invalid
parameter is entered, the Parameter out of range
message will be returned.