Trigger[1..n][:sequence]:source – EXFO IQS-1600 High-Speed Power Meter for IQS-500/600 User Manual
Page 228

SCPI Command Reference
220
IQS-1600
Product-Specific Commands—Description
:TRIGger[1..n][:SEQuence]:SOURce?
Description
This query returns the trigger condition's source.
Syntax
:TRIGger[1..n][:SEQuence]:SOURce?
Parameter(s)
None
Response Syntax
<TriggerSource>
Response(s)
TriggerSource:
The response data syntax for <TriggerSource> is
defined as a <CHARACTER RESPONSE DATA>
element.
The current trigger source, where:
EXTernal - is the external trigger
INTernal1 - is channel 1
INTernal2 - is channel 2 (IQS-1620 or IQS-1640)
INTernal3 - is channel 3 (IQS-1640) or virtual
channel 1(IQS-1620)
INTernal4 - is channel 4 (IQS-1640)
INTernal5 - is virtual channel 1 (IQS-1640)
INTernal6 - is virtual channel 2 (IQS-1640)
Example(s)
TRIG:SEQ:SOUR?
See Also
TRIGger[:SEQuence]:SOURce
TRIGger[:SEQuence]:LEVel?
TRIGger:POSition
TRIGger:POSition:CATalog?
TRIGger[:SEQuence]:STATe