Using the External Trigger and Analog Outputs
84
IQS-1600
Analog Output (IQS-1610, IQS-1620)
The following is an illustration of typical uses of the analog output.
The user interface can be used to determine the operating power levels
and to optimize the analog output range. Choose the optimum power
range for your application. The table below provides the voltage vs. power
information for each available manual range.
Manual
Range
Power vs. Voltage
a
a.
At 1310 nm.
LR1, HR1
2.15 V ± 25 %
b
at 18.30 dBm
c
b.
Large uncertainty is due to detector responsivity at 1310 nm.
c.
Not guaranteed due to detector saturation.
LR2, HR2
2.15 V ± 25 %
b
at 9.90 dBm
LR3, HR3
2.15 V ± 25 %
b
at −10.09 dBm
LR4, HR4
2.15 V ± 25 %
b
at −29.87 dBm
LR5
2.15 V ± 25 %
b
at −49.61 dBm or at 44.94 dBm for IQS-16XXW models
LR6
2.15 V ± 25 %
b
at −69.61 dBm or at 64.10 dBm for IQS-16XXW models
IQS-1600
Analog output
Digital oscilloscope
2-Ch Power Meter
IQS-1600
EXT.
TRIG
ANALOG OUT
C1
C2
IQS-1600
Analog output
Data logger
2-Ch Power Meter
IQS-1600
EXT.
TRIG
ANALOG OUT
C1
C2