Trigger[1..n][:sequence]:slope – EXFO IQS-1600 High-Speed Power Meter for IQS-500/600 User Manual
Page 225

SCPI Command Reference
High Speed Power Meter
217
Product-Specific Commands—Description
:TRIGger[1..n][:SEQuence]:SLOPe
Description
This command sets the edge status of the trigger
condition. It defines whether acquisitions will be
triggered on positive or negative transitions.
Syntax
:TRIGger[1..n][:SEQuence]:SLOPe<wsp>NEGati
ve|POSitive
Parameter(s)
Slope:
The program data syntax for the first parameter is
defined as a <CHARACTER PROGRAM DATA>
element. The allowed <CHARACTER PROGRAM
DATA> elements for this parameter are:
NEGative|POSitive.
This parameter sets the trigger edge.
NEGative - represents a falling edge trigger.
POSitive - represents a rising edge trigger.
Example(s)
TRIG:SEQ:SLOP POS
See Also
TRIGger{:SEQuence]:SLOPe?
TRIGger[:SEQuence]:LEVel
TRIGger:POSition
TRIGger:POSition:CATalog?
TRIGger[:SEQuence]:STATe
TRIGger[:SEQuence]:SOURce