Iq pneumatic press error messages – Dukane Ultrasonic Generator 403-575-00 User Manual

Page 90

Advertising
background image

Table 6 - II

Error Messages, Pneumatic-related

iQ

Pneumatic Press Error Messages

Error Message

Description

Cause(s)

Possible Solution(s)

Overload (average) at xxx msecs. Power above generator rating.

Power from ultrasonic stack sent to the part exceeds power rating of generator

.

1. Excessive pressure applied to the part. 2. Booster size is too large.

1. Decrease pressure applied to part. 2. Reduce size of booster

.

Overload (peak) at xxx msecs. Check stack.

1. Component failure in the stack. 2. Improper stack assembly

.

3. Incorrect

Advanced Hardware settings.

1. Check functionality of stack components. 2. Disassemble and reassemble stack. 3. Check the generator's

Advanced Hardware

settings.

Over temperature fault!

Generator's internal temperature ex

-

ceeds 158º F (70º Celsius).

1. W

elder duty cycle is too large.

2. W

eld/hold times or forces are too

large. 3.

Thruster fan not working.

4. Servo actuator or amplifier is dam

-

aged.

1. Reduce cycle rate. 2. Change setup to reduce weld/hold forces or decrease weld duration. 3. Check if fan is obstructed; contact DUKANE service for replacement. 4. Contact DUKANE service.

Overload (frequency) at xxx msecs. Check stack, cable or validate ramp up time.

Generator is unable to lock to the reso

-

nant frequency of the ultrasonic stack.

1. Improper stack assembly

.

2. Defective ultrasound cable. 3. Ramp up time is too short.

1. Disassemble and reassemble stack. 2. Replace ultrasound cable. 3. Increase ramp up time.

Overload (frequency) at xxx msecs. Check for coupling between stack and part.

Generator correctly locks to the resonant frequency of the ultrasonic stack, but loses the lock at a point later in the weld cycle.

Horn comes in contact with the part nest or anvil.

Make adjustments to thruster and part nest.

Power fault. Check

AC line.

AC line voltage drops below 190 V

AC on

240 V

AC systems, and below 95 V

AC on

120 V

AC systems.

Fluctuations in

AC power

.

Check the incoming

AC line power

.

Page 84

iQ Series

Ultrasonic Generator/Power Supply ES User’s Manual

Dukane Manual Part No. 403-575-00