Chapter 3: Getting Started

Product Service 1-800-522-7658

16

Installation of Gloves to the Glove Ports

With the thumbs up and in a right/left orientation, secure the gloves in place
on the glove ports by stretching the beaded glove cuff into the glove port
groove nearest the window. Install the separate 8" diameter O-ring over the
gloves, into the outer groove of the glove port surface. Stainless steel band
clamps are provided for securing the separate O-ring into the glove port
groove. Replacement gloves and parts are listed in Chapter 7.

Figure 3-3

Glove Installation

Connecting the Electrical Supply Source to the
Glove Box

Connect the power cords supplied to both IEC electrical supply plugs located
on the back of the glove box. For doublewide models, connect one extra
power cord for the centrally located duplex on the left side. 100-115V
models are rated at 12 amps for each power cord. 230V models are rated at
12 amps for one power cord and 6 amps for the other power cord to power
the exterior vacuum pump outlet. The maximum circuit load for the
interior electrical duplex is 10 amps.
The exterior vacuum pump receptacle
is rated at 8 amps for 100-115V models and 4.5 amps for 230V models. On
auto pressure control and combination models, a foot switch for pressure
control is supplied and is connected (3 position connector) to the back of the
glove box as marked. See Figure 3-2 for all electrical connections. See
Figures 1-1 and 1-3 for foot switch.